FDA-Device2024-01-24Class II

Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351

Randox Laboratories Ltd.
Hazard

Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37¿C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples

Sold states
US Nationwide including Puerto Rico.
Affected count
108 kits
Manufactured in
55 Diamond Road, N/A, Crumlin (North), N/A, Ireland
Products
Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0753-2024

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief