FDA-Device2013-02-06Class II
Ortho Clinical Diagnostics VITROS¿ Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS¿ Vapor Adsorption Cartridges are used to remove vapor contaminants that may potentially interfere with Signal Reagent used with MicroWell Assays on the VITROS¿ 3600, VITROS¿ 5600, and VITROS¿ ECi/ECiQ Systems.
Ortho-Clinical Diagnostics
Hazard
Some VITROS¿ Chemistry Products Vapor Adsorption Cartridge CAT No. 6800100 cartons incorrectly contain a Particulate Cartridge Part No. J19612. In some instances, the Particulate Cartridge has been inadvertently installed on the VITROS¿ System instead of the intended VITROS¿ Vapor Adsorption Cartridge which may bias results.
Sold states
Worldwide Distribution, including Nationwide (USA) and the countries of AU, GM, FR, IT, SP, BR, BM, AR, CA, JA, MX, UK, CH, CL, CO, IN, EC, PE, PY, PA, PT, NZ, SG, VE and UY.
Affected count
Domestic: 15,891; Foreign: 5237
Manufactured in
100 Indigo Creek Dr, Rochester, NY, United States
Products
Ortho Clinical Diagnostics VITROS¿ Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS¿ Vapor Adsorption Cartridges are used to remove vapor contaminants that may potentially interfere with Signal Reagent used with MicroWell Assays on the VITROS¿ 3600, VITROS¿ 5600, and VITROS¿ ECi/ECiQ Systems.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0755-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief