FDA-Device2021-01-06Class II
Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for clinical use Part Number: 05641446001
Roche Diagnostics Operations, Inc.
Hazard
Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.
Sold states
US Nationwide distribution.
Affected count
7 units
Manufactured in
9115 Hague Rd, N/A, Indianapolis, IN, United States
Products
Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for clinical use Part Number: 05641446001
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0755-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.