FDA-Device2016-12-28Class III

ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T NON-ECHO S PG, Item Number H787065970655, Catalog No.REF 06597065, 2) 4 F Stiffened Micro-Introducer Kit, STF 4F M.I. KIT 45CM NT/T NON-ECHO S PG, Item Number H787065970695, Catalog No.REF 06597069 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick

Angiodynamics, Inc.
Hazard

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

Sold states
Nationwide Distribution to AL, AR, CA, FL, GA, IL, KY, LA, MI, MN, MO, NC, OH, OK, SC, NY, ME, PA, MD, VA, WA, WI, WV, TN, and TX
Affected count
145 boxes
Manufactured in
603 Queensbury Ave, N/A, Queensbury, NY, United States
Products
ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T NON-ECHO S PG, Item Number H787065970655, Catalog No.REF 06597065, 2) 4 F Stiffened Micro-Introducer Kit, STF 4F M.I. KIT 45CM NT/T NON-ECHO S PG, Item Number H787065970695, Catalog No.REF 06597069 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0776-2017