FDA-Device2023-01-11Class II

Guardian iOS app (MMT-8200) and Guardian Android app (MMT-8201), part of the Guardian 4 system

Medtronic MiniMed
Hazard

An app, part of a continuous glucose monitoring system, for use with smartphone devices may automatically log out from CareLink, then the app is not able to upload data. When logged out, linked care partners will not receive SMS notifications (could result in hypoglycemia or hyperglycemia), and sensor glucose values will also not be sent to the InPen app.

Sold states
OUS: Austria, Belgium, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Luxembourg, Moldova, Republic of, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom
Affected count
3235
Manufactured in
18000 Devonshire St, N/A, Northridge, CA, United States
Products
Guardian iOS app (MMT-8200) and Guardian Android app (MMT-8201), part of the Guardian 4 system

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0776-2023