FDA-Device2023-01-11Class II

VERIFY Steam Test Pack (20 Tests and 20 controls) Item Number: S3069 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes

Steris Corporation
Hazard

Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.

Sold states
Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA
Affected count
105 boxes
Manufactured in
5960 Heisley Rd, N/A, Mentor, OH, United States
Products
VERIFY Steam Test Pack (20 Tests and 20 controls) Item Number: S3069 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0781-2023

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.