FDA-Device2020-01-15Class II

Cios Alpha, Model Number 10308191; mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical application

Siemens Medical Solutions USA, Inc
Hazard

During a routine inspection of a system, a crack on the C-arm holder was discovered. In the event the C-arm holder has a crack, the stability of the C-arm may be compromised and could lead to deformations of the C-arm. In very unlikely cases, an extreme impact to the C-arm such as heavy collision with a wall or table, may lead to a detachment of the C-arm and could potentially result in injury to patients and staff.

Sold states
US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
Affected count
190 (updated total of 202 as of 1/20/20)
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Cios Alpha, Model Number 10308191; mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical application

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0782-2020

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.