FDA-Device2014-01-29Class II

ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH

Philips Medical Systems (Cleveland) Inc
Hazard

The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.

Sold states
Worldwide Distribution: US (nationwide) including states of: AL, AR, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TN, TX, VA, VT, WA, WI & WV; and Internationally to: Austria, Belgium, Canada, Denmark, Germany, Great Britain, Italy, Netherlands, Norway, Pakistan, Russia, Singapore & South Korea.
Affected count
255 units
Manufactured in
595 Miner Rd, Cleveland, OH, United States
Products
ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0785-2014