FDA-Device2019-02-06Class II
SUCTION / ANTICOAGULATION ASSEMBLY (a) BTC93 (b) BTC98 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.
Hazard
Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.
Sold states
Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Japan, and Sweden.
Affected count
7587 units
Manufactured in
7611 Northland Dr N, N/A, Brooklyn Park, MN, United States
Products
SUCTION / ANTICOAGULATION ASSEMBLY (a) BTC93 (b) BTC98 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0786-2019More Medtronic Perfusion Systems recalls
- FDA-Device2026-07-22Medtronic Bio-Medicus Adult Cannula Kit recalled for loose foreign material with blood
- FDA-Device2026-07-22Medtronic DLP One-Piece Pediatric Arterial Cannula recalled for loose foreign material with blood
- FDA-Device2026-07-01Medtronic Octopus Evolution AS Tissue Stabilizer recalled for reversed tubing assembly
- FDA-Device2026-07-01Medtronic Octopus 4 Tissue Stabilizer recalled for reversed tubing assembly
- FDA-Device2026-07-01Medtronic Octopus Evolution Tissue Stabilizer recalled for incorrect tubing assembly
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief