FDA-Device2013-02-13Class II

Oxoid Antimicrobial Susceptibility Test Discs, Aztreonam 30 ug, IVD, REF CT0264B, Oxoid Ltd., UK, oxoid@thrmofisher.com (Both Remel & Oxoid, Ltd are a part of Thermo Fisher Scientific) An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure, by a disc-agar diffusion technique or a disc-broth elution technique, the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial a

Remel Inc
Hazard

The firm is recalling the product due to some cartridges may contain individual discs that are not impregnated with antibiotic. The firm notified their customers of the recall with a notification letter and requested a confirmation fax back of receipt.

Sold states
Nationwide Distribution including WI, MO, MA, CT, AL, IL, WA, CA, RI, MD, NJ, NC and TX.
Affected count
30
Manufactured in
12076 Santa Fe Trail Dr, N/A, Lenexa, KS, United States
Products
Oxoid Antimicrobial Susceptibility Test Discs, Aztreonam 30 ug, IVD, REF CT0264B, Oxoid Ltd., UK, oxoid@thrmofisher.com (Both Remel & Oxoid, Ltd are a part of Thermo Fisher Scientific) An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure, by a disc-agar diffusion technique or a disc-broth elution technique, the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents (Aztreonam).

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0790-2013