FDA-Device2016-03-02Class I
Single Lumen Central Venous Catheter Sets and Trays The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.
Cook Inc.
Hazard
A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.
Sold states
Worldwide Distribution - US Nationwide including DC and PR including the countries of Australia, Germany, China, Chile, Dominican Republic, Panama, Israel, Brazil, Canada
Affected count
1341
Manufactured in
750 N Daniels Way, N/A, Bloomington, IN, United States
Products
Single Lumen Central Venous Catheter Sets and Trays The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0799-2016Don't want to check this manually?