FDA-Device2016-02-24Class II
Exactech Acapella One Interbody Fusion Device Parallel. Orthopedic Spinal Surgery Device. Product Usage: The Acapella One Cervical Spacer System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease at one disc level from C2-T1. Is to be used with autogenous bone graft and supplemental internal fixation and is implanted via an open anterior approach.
Exactech, Inc.
Hazard
Incorrect external and internal packaging labels. Exactech has identified nineteen (19) Acapella One Interbody Fusion Device Lordotic (05-090-01-1610) were packaged as Acapella One Interbody Fusion Device Parallel (05-090-02-1610)
Sold states
US distribution in the state of FL and PA.
Affected count
19 pieces
Manufactured in
2320 NW 66th Ct, Gainesville, FL, United States
Products
Exactech Acapella One Interbody Fusion Device Parallel. Orthopedic Spinal Surgery Device. Product Usage: The Acapella One Cervical Spacer System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease at one disc level from C2-T1. Is to be used with autogenous bone graft and supplemental internal fixation and is implanted via an open anterior approach.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0809-2016Don't want to check this manually?