FDA-Device2025-01-15Class II
Tec 6 Plus Anesthesia Vaporizer, Reference Numbers 1107-9601-000, 1107-9602-000, 1107-9603-000, 1107-9604-000, 1107-9610-000, 1107-9612-000, 1107-9616-000, 1107-9617-000, 2066835-001, 2069767-001, 2078800-001, 2091008-001, 2091009-001, M1098876, M1145610, M1220023, M1220854, M1221495, and M1233992
GE Medical Systems, LLC
Hazard
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Sold states
US Nationwide. Global Distribution.
Affected count
1573 units
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Products
Tec 6 Plus Anesthesia Vaporizer, Reference Numbers 1107-9601-000, 1107-9602-000, 1107-9603-000, 1107-9604-000, 1107-9610-000, 1107-9612-000, 1107-9616-000, 1107-9617-000, 2066835-001, 2069767-001, 2078800-001, 2091008-001, 2091009-001, M1098876, M1145610, M1220023, M1220854, M1221495, and M1233992
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0814-2025Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.