FDA-Device2014-01-29Class II

ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br¿nsh¿j, Denmark A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.

Radiometer America Inc
Hazard

RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.

Sold states
Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI & WV and Internationally to Australia, Austria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hungary, Italy, Japan, Netherlands, Norway, Poland, Russia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey & United Kingdom.
Affected count
Total 3414 (US 524, CA 31, ROW 2859)
Manufactured in
810 Sharon Dr, Westlake, OH, United States
Products
ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br¿nsh¿j, Denmark A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0816-2014