FDA-Device2014-12-24Class II
VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and 3.5.0.
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Hazard
Software Anomaly. The VERO MHI-TM2000 Operator Console may provide Patient Positioning System (ExacTrac) with Image Angle Information Used for the 1st Port Image, for a subsequent port image fusion. As the result, ExacTrac may display a Digitally Reconstructed Radiograph (DRR) image taken from the angle for the 1st port image fusion, rather than the one taken from the angle for the intended port
Sold states
US Distribution to the states of NY, FL, TX and OH.
Affected count
Domestic: 4 units
Manufactured in
6-22, 4-CHOME, KAN-ON-SHIN-MACHI, NISHI-KU, HIROSHIMA, N/A, Japan
Products
VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and 3.5.0.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0821-2015Don't want to check this manually?