FDA-Device2020-01-29Class II
Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3
Isopure Corp
Hazard
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.
Sold states
The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.
Affected count
15
Manufactured in
11851 Plantside Dr, Louisville, KY, United States
Products
Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0821-2020Don't want to check this manually?