FDA-Device2025-01-22Class I

Impella RP with SmartAssist; Product Number: 0046-0035;

Abiomed, Inc.
Hazard

Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.

Sold states
Domestic: Nationwide Distribution International: Canada
Affected count
1,560 units
Manufactured in
24 Cherry Hill Dr, Danvers, MA, United States
Products
Impella RP with SmartAssist; Product Number: 0046-0035;

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0829-2025