FDA-Device2024-02-07Class II
Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) C-SECTION PACK, kit number FDCS01AA; b) C-SECTION PACK, kit number FDCS01AB; c) C-SECTION STERILE PACK, kit number HGCS68J; d) C-SECTION PACK, kit number HICS50H; e) C-SECTION PPS, kit number LLCI66; f) C-Section drape Pack, kit number RBCS12R; g) C-Section, kit number THCS49B; h) C-SECTION PACK (PS 020632), kit number UICS50AG;
American Contract Systems, Inc.
Hazard
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Sold states
US Nationwide distribution.
Affected count
1818 units
Manufactured in
7802 E Telecom Pkwy, Temple Terrace, FL, United States
Products
Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) C-SECTION PACK, kit number FDCS01AA; b) C-SECTION PACK, kit number FDCS01AB; c) C-SECTION STERILE PACK, kit number HGCS68J; d) C-SECTION PACK, kit number HICS50H; e) C-SECTION PPS, kit number LLCI66; f) C-Section drape Pack, kit number RBCS12R; g) C-Section, kit number THCS49B; h) C-SECTION PACK (PS 020632), kit number UICS50AG;
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0830-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief