Impella RP Flex with SmartAssist; Product Number: 1000323;
Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0830-2025More Abiomed, Inc. recalls
- FDA-Device2026-06-03Abiomed Impella Controller recalled for alarm failures and cassette misalignment
- FDA-Device2026-06-03Abiomed Impella Controller recalled for hardware defects affecting motor control and device function
- FDA-Device2026-04-08Abiomed Impella Controller recalled for delayed alarm display with Impella 5.5 pumps
- FDA-Device2026-04-08Abiomed Impella Pump Set purge cassettes recalled for increased leak risk