FDA-Device2014-02-05Class II
Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.
DeRoyal Industries Inc
Hazard
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
Sold states
Nationwide Distribution.
Affected count
738 units
Manufactured in
1595 Highway 33 S, New Tazewell, TN, United States
Products
Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0834-2014Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.