FDA-Device2014-12-31Class II
adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
Ion Beam Applications S.A.
Hazard
Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du
Sold states
US Distribution in PA only.
Affected count
1
Manufactured in
Chemin du Cyclotron, 3, Louvain La Neuve, Belgium
Products
adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0836-2015Don't want to check this manually?