FDA-Device2018-03-14Class II

Brilliance Big Bore Radiology CT with 4.2.0 software version. Model #' 728244 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.

Philips Medical Systems (Cleveland) Inc
Hazard

Software issues including: Software errors that may result in Extended Field of View check message not displayed when Field of View is set >600 mm for a paused scan; Request for change to automatically turn on the center x/y 0 setting and the disable result rotation setting when turning on the CT simulation Exam Card; Software errors that may result in Tumor LOC allowing POI (Point of Interest) locking when a non-isocenter slice is displayed; Software errors that may result in partial set of images or inability to generate CT images.

Sold states
Worldwide Distribution - US Nationwide
Affected count
119
Manufactured in
595 Miner Rd, N/A, Cleveland, OH, United States
Products
Brilliance Big Bore Radiology CT with 4.2.0 software version. Model #' 728244 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0839-2018