FDA-Device2022-04-06Class II

Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system

Siemens Medical Solutions USA, Inc
Hazard

Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If the user activates a longitudinal or transversal movement at the time of the error, the system controls may initiate a correction of the tube stand position, resulting in an unintended movement. If the described error occurs, a potential for collision risk for objects or persons in the vicinity of the moving stand exists. This may result in minor to serious injuries to persons.

Sold states
US Nationwide.
Affected count
70 Systems
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0841-2022