FDA-Device2020-01-29Class II
Instrumentarium Dental Express SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Palodex Group Oy
Hazard
Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.
Sold states
US: WI and PA OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, REPUBLIC OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM
Affected count
4,563 Total Globally (all products)
Manufactured in
Nahkelantie 160, Tuusula, Finland
Products
Instrumentarium Dental Express SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0842-2020Don't want to check this manually?