FDA-Device2021-01-20Class II
CareLink Personal Software Therapy Management Software Tool for Diabetes CareLink System Therapy Management Software Tool for Diabetes Software Uploader versions: 3.1.5.000 3.1.2.000
Medtronic Inc.
Hazard
Incomplete basal rate information may be transmitted leading to a potential for inappropriate change in therapy. During data upload, radio frequency (RF) interference can occur at any point from surrounding devices or equipment that may affect the data transfer between the insulin pump and the CareLink Personal or CareLink System Software.
Sold states
U.S. Nationwide O.U.S. Worldwide Refer to Associated Documents and uploaded Consignee Lists/Spreadsheets. -Australia New Zealand -Canada -EMEA -Greater China -India -Japan -Latin America -South Korea -Southeast Asia -US
Affected count
46,461 affected systems
Manufactured in
18000 Devonshire St, N/A, Northridge, CA, United States
Products
CareLink Personal Software Therapy Management Software Tool for Diabetes CareLink System Therapy Management Software Tool for Diabetes Software Uploader versions: 3.1.5.000 3.1.2.000