FDA-Device2014-02-05Class II
InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended to induce, maintain and reverse mild hypothermia, to achieve and/or maintain normothermia, and for use in fever reduction. K080908
Philips Healthcare
Hazard
Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.
Sold states
Worldwide Distribution - US (Nationwide) and Internationally to: Austria, France, Germany The Netherlands, Spain, Sweden, and Switzerland.
Affected count
50
Manufactured in
6740 Top Gun St, San Diego, CA, United States
Products
InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended to induce, maintain and reverse mild hypothermia, to achieve and/or maintain normothermia, and for use in fever reduction. K080908
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0845-2014Don't want to check this manually?