FDA-Device2014-02-05Class II

Kimberly-Clark RadiOpaque Radiofrequency Cannula, Gauge 22, Length 100 mm, Active Tip 10 mm, Distributed in the US by Kimberly-Clark,Product code PMF22-100-10 The product is used to create lesions in nervous tissue when used with the Kimberly-Clark radiofrequency generator and probes.

Kimberly-Clark Corporation
Hazard

Product may contain a cannula with an active tip length of only 5mm rather than 10mm as indicated on the label.

Sold states
US Distribution in states of: AZ, CA, DE, FL, LA, MD, MA, NJ, NY, OH, PA and TX.
Affected count
55 cases (10 units per case = 550 units)
Manufactured in
1400 Holcomb Bridge Rd, Roswell, GA, United States
Products
Kimberly-Clark RadiOpaque Radiofrequency Cannula, Gauge 22, Length 100 mm, Active Tip 10 mm, Distributed in the US by Kimberly-Clark,Product code PMF22-100-10 The product is used to create lesions in nervous tissue when used with the Kimberly-Clark radiofrequency generator and probes.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0850-2014