FDA-Device2018-03-14Class II

Philips Network Firewall (Cisco ASA 5506), Model # 867098

Philips Electronics North America Corporation
Hazard

Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX. The defect involves the clock signal component within the firewall. This component has a high probability of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable. This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot.

Sold states
US Nationwide
Affected count
332 units
Manufactured in
3000 Minuteman Rd, N/A, Andover, MA, United States
Products
Philips Network Firewall (Cisco ASA 5506), Model # 867098

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0850-2018