FDA-Device2021-01-20Class II

GE NM/CT 870 CZT system Model Number H3906CW - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.

GE Healthcare, LLC
Hazard

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Sold states
Worldwide
Affected count
12 units
Manufactured in
3000 N Grandview Blvd, N/A, Waukesha, WI, United States
Products
GE NM/CT 870 CZT system Model Number H3906CW - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0855-2021

Don't want to check this manually?