FDA-Device2021-01-20Class II

Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage: intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, usin

GE Healthcare, LLC
Hazard

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Sold states
Worldwide
Affected count
52 untits
Manufactured in
3000 N Grandview Blvd, N/A, Waukesha, WI, United States
Products
Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage: intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0857-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief