VITROS CRP Slides: 1) VITROS¿ Chemistry Products CRP Slides (250 slides per sales unit), Catalog # 1926740 and 2) , VITROS¿ Chemistry Products CRP Slides (90 slides per sales unit), Catalog # 1926740
CRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result. The IFUs do not provide specific instructions for how to account for endogenous CRP in the diluent when calculating the sample result. Orthos diluents used to dilute samples containing CRP are manufactured using human plasma based fluids that contain endogenous CRP. Without accounting for the CRP in the diluent and removing its contribution from the result, the reported result will include the CRP present in both the sample and diluent, possibly leading to positively biased results.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0864-2017Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief