FDA-Device2014-02-05Class II

T3F, Free T3 Assay, REF/Catalog Number L2KF32 (200 tests), Siemens Material Number (SMN) 10381675, and REF/Catalog Number L2KF36 (600 Tests), SMN 10381682; an IVD Immunoassay kit for use with the IMMULITE 2000/IMMULITE 200 XPi Analyzers --- Origin: UK Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL55 4EL UK. For the quantitative measurement of Free T3 in serum, as an aid in the clinical assessment of thyroid status.

Siemens Healthcare Diagnostics
Hazard

Siemens Healthcare Diagnostics confirmed customer complaints regarding an increase in the number of euthyroid patients (those with normal function of the thyroid) demonstrating values above the recommended normal range as published in the Instructions For Use for the IMMULITE/IMMULITE 1000 and/or IMMULITE 2000/IMMULITE 2000 XPi for several lots of Free T3 kits. A positive bias in quality control results was also observed, but the values may remain within the established ranges.

Sold states
Worldwide Distribution-USA (nationwide) and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bulgaria, Bosnia-Herzegovina, Brazil, Belarus, Canary Islands, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Great Britain, Greece, Guatemala, Honduras, Hungary, Iceland, India, Indonesia, Ireland, Israel, Iran, Italy, Jordan, Republic of South Korea, Kuwait, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Macedonia, Mexico, Malaysia, Malta, Nepal, Netherlands, Norway, Panama, Peru, Philippines, Pakistan, Paraguay, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sri Lanka, South Africa, Switzerland, Turkey, Thailand, Taiwan, Uruguay, Usbekistan, Venezuela, Vietnam, and United Arab Emirates.
Affected count
4695 kits in total
Manufactured in
511 Benedict Ave, Tarrytown, NY, United States
Products
T3F, Free T3 Assay, REF/Catalog Number L2KF32 (200 tests), Siemens Material Number (SMN) 10381675, and REF/Catalog Number L2KF36 (600 Tests), SMN 10381682; an IVD Immunoassay kit for use with the IMMULITE 2000/IMMULITE 200 XPi Analyzers --- Origin: UK Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL55 4EL UK. For the quantitative measurement of Free T3 in serum, as an aid in the clinical assessment of thyroid status.