FDA-Device2020-02-05Class II

EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light source emitting incoherent non-laser light from the Deep UV through the visible wavelengths.

Energetiq Technology Inc
Hazard

On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted that a safety interlock within the lamp does not function in a reliable manner to shut down the system if the laser collimator is removed from the lamp while the system is operating. This is a potential safety hazard for anyone working on or nearby the lamp.

Sold states
Worldwide Distribution.
Affected count
15
Manufactured in
7 Constitution Way, N/A, Woburn, MA, United States
Products
EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light source emitting incoherent non-laser light from the Deep UV through the visible wavelengths.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0870-2020

Don't want to check this manually?