FDA-Device2013-02-27Class II
Liko Overhead Rail System, Ceiling Mounted Rail Systems for all Liko Stationary Lifts including Likorall, Multirall and Masterlift Systems Product Usage - The Overhead Rail System allows overhead lifts to be mounted to the ceiling allowing versatility for use in lifting situations. It can be used with different Liko ceiling lifts.
Hill-Rom, Inc.
Hazard
Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity (swimming pool) areas. Analysis of the incidents suggest that the components (hardware) used to secure the rail to the ceiling had corroded and may be unsuitable for environments with an atmosphere con
Sold states
Worldwide Distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and CANADA LOCATIONS: AB, BC, MB, NB, NS, ON, and SK.
Affected count
3816 in total
Manufactured in
1069 State Route 46 East, N/A, Batesville, IN, United States
Products
Liko Overhead Rail System, Ceiling Mounted Rail Systems for all Liko Stationary Lifts including Likorall, Multirall and Masterlift Systems Product Usage - The Overhead Rail System allows overhead lifts to be mounted to the ceiling allowing versatility for use in lifting situations. It can be used with different Liko ceiling lifts.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0872-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.