FDA-Device2017-01-04Class II

NeuViz 64 Multi-slice CT Scanner Systems

Neusoft Medical Systems Co., Ltd.
Hazard

The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the interruption error before scanning can resume.

Sold states
US Distribution
Affected count
23 units
Manufactured in
NO. 16 Shiji Road, Hunnan Industrial Area, Shenyang, N/A, China
Products
NeuViz 64 Multi-slice CT Scanner Systems

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0875-2017

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief