FDA-Device2021-01-27Class II
Reliance Vision Single-Chamber Washer/Disinfector - Product Usage: intended for use in the cleaning and intermediate level disinfection of soiled reusable utensils, trays, glassware, bedpans and urinals, rubber and plastic goods, simple hard-surfaced rigid surgical instruments, such as forceps and clamps, theatre shoes, and other similar and related articles found in healthcare facilities.
Steris Corporation
Hazard
The electrical contactor component present in the drying chamber of the Reliance Vision Single-Chamber Washer/Disinfector may malfunction, potentially leading to the heating elements in the drying chamber to overheat, eventually resulting in smoke and/or fire.
Sold states
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Bulgaria, Canada, Chile, China, Ecuador, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of Malaysia, Mexico, New Zealand, Panama, Portugal, Russian Federation, Saudi Arabia, Singapore, Spain, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela.
Affected count
3405
Manufactured in
5960 Heisley Rd, N/A, Mentor, OH, United States
Products
Reliance Vision Single-Chamber Washer/Disinfector - Product Usage: intended for use in the cleaning and intermediate level disinfection of soiled reusable utensils, trays, glassware, bedpans and urinals, rubber and plastic goods, simple hard-surfaced rigid surgical instruments, such as forceps and clamps, theatre shoes, and other similar and related articles found in healthcare facilities.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0876-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief