FDA-Device2023-01-18Class I
MAHURKAR 13.5Fr High Flow Dual Lumen Acute Dialysis Catheter, Sterile, Single Use, Rx Only: 13.5cm, Straight Extensions, Kit - Model No. 8888135131; 13.5 cm, Curved Extensions, Kit - Model No. 8888135132; 13.5cm, Pre-Curved, Kit - Model No. 8888135133; 13.5cm, Straight Extensions, IC Tray, Model No. 8888135134; 16cm, Straight Extensions, Kit - Model No. 8888135161; 16cm, Curved Extensions, Kit - Model No. 8888135162; 16 cm, Pre-Curved, Kit - Model No. 8888135163; 16cm, Straig
Covidien LP
Hazard
There is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
Sold states
Worldwide distribution. US nationwide including Guam, USVI, and Puerto Rico; Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Polynesia, Germany, Greece, Hungary, Iceland, Italy, Jamaica, Kenya, Malaysia, Malta, Mauritius, Mexico, Nicaragua, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Uganda, United Arab Emirates, and United Kingdom.
Affected count
359,730 (23,003 US; 336,727 OUS)
Manufactured in
15 Hampshire St, Mansfield, MA, United States
Products
MAHURKAR 13.5Fr High Flow Dual Lumen Acute Dialysis Catheter, Sterile, Single Use, Rx Only: 13.5cm, Straight Extensions, Kit - Model No. 8888135131; 13.5 cm, Curved Extensions, Kit - Model No. 8888135132; 13.5cm, Pre-Curved, Kit - Model No. 8888135133; 13.5cm, Straight Extensions, IC Tray, Model No. 8888135134; 16cm, Straight Extensions, Kit - Model No. 8888135161; 16cm, Curved Extensions, Kit - Model No. 8888135162; 16 cm, Pre-Curved, Kit - Model No. 8888135163; 16cm, Straight Extensions, IC Tray - Model No. 8888135164; 19.5cm, Straight Extensions, Kit - Model No. 8888135191; 19.5cm, Curved Extensions, Kit - Model No. 8888135192; 19.5cm, Pre-Curved, Kit - Model No. 8888135193; 19.5cm, Straight Extensions, IC Tray - Model No. 8888135194; 19.5cm, Curved Extensions, PASS Tray - Model No. 8888135198; 24cm, Straight Extensions, Kit - Model No. 8888135241; 24cm, Curved Extensions, Kit - Model No. 8888135242; 24cm, Pre-Curved, Kit - Model No. 8888135243
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0877-2023Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief