FDA-Device2021-01-27Class II
Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25
Abbott Point Of Care Inc.
Hazard
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
Sold states
Nationwide
Affected count
2,820,000 cartridges US and 2,273,453 catridges OUS
Manufactured in
400 College Rd E, N/A, Princeton, NJ, United States
Products
Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0880-2021Don't want to check this manually?