FDA-Device2013-03-06Class II

Aesculap Bipolar Generator Foot Pedal, Catalog No. GK226, for electrosurgical cutting and coagulation.

Aesculap, Inc.
Hazard

The bipolar energy did not stop after release of the foot pedal.

Sold states
USA Nationwide Distribution including the states of: AR, CA, FL, IN, IA, KS, MD, MN, MO, OR, TN, TX, VA and WA.
Affected count
59
Manufactured in
3773 Corporate Pkwy, N/A, Center Valley, PA, United States
Products
Aesculap Bipolar Generator Foot Pedal, Catalog No. GK226, for electrosurgical cutting and coagulation.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0887-2013