FDA-Device2013-03-06Class II
Aesculap Bipolar Generator Foot Pedal, Catalog No. GK226, for electrosurgical cutting and coagulation.
Aesculap, Inc.
Hazard
The bipolar energy did not stop after release of the foot pedal.
Sold states
USA Nationwide Distribution including the states of: AR, CA, FL, IN, IA, KS, MD, MN, MO, OR, TN, TX, VA and WA.
Affected count
59
Manufactured in
3773 Corporate Pkwy, N/A, Center Valley, PA, United States
Products
Aesculap Bipolar Generator Foot Pedal, Catalog No. GK226, for electrosurgical cutting and coagulation.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0887-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief