FDA-Device2018-03-14Class II
LASEREDGE Knives, 1.1MM ARROW KNIFE ANGLED (6/BOX), REF/PRODUCT CODE E7592, STERILE, Rx Only Product Usage: The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures.
Bausch & Lomb Inc Irb
Hazard
The complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing.
Sold states
US Nationwide Distribution
Affected count
1,236,749 knives in total
Manufactured in
1400 Goodman St N, N/A, Rochester, NY, United States
Products
LASEREDGE Knives, 1.1MM ARROW KNIFE ANGLED (6/BOX), REF/PRODUCT CODE E7592, STERILE, Rx Only Product Usage: The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures.