FDA-Device2022-04-13Class II

ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.

Abbott Laboratories
Hazard

This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).

Sold states
Worldwide Distribution: US (nationwide) and OUS (foreign) including countries of: AUSTRIA, BELGIUM, BOSNIA AND HERZEGOVI, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, GREECE, IRAQ, ITALY, KAZAKHSTAN, LITHUANIA, LUXEMBOURG, MACEDONIA, MOROCCO, PANAMA, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, and TURKEY
Affected count
402 kits
Manufactured in
100 Abbott Park Rd Bldg Ap8b, Abbott Park, IL, United States
Products
ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0891-2022