FDA-Device2016-12-28Class II
ev3, PowerCross .018 OTW PTA Dilatation Catheter, Sterile EO. The PowerCross .018 OTW PTA Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.
Medtronic Inc
Hazard
Medtronic identified an issue with specific models and lots of the PowerCross .018 Over the Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Dilatation Catheter. During an unrelated manufacturing inspection of scrapped products, an out of specification condition was observed. The catheter outer shafts were pinched and deformed at the proximal balloon bond. Bench testing of affected devices showed that this resulted in either the inability to inflate the device, a slow deflation, or the inability to deflate the device post expansion.
Sold states
NY, NC, WA
Affected count
315
Manufactured in
710 Medtronic Pkwy, NE Mailstop LS245, Minneapolis, MN, United States
Products
ev3, PowerCross .018 OTW PTA Dilatation Catheter, Sterile EO. The PowerCross .018 OTW PTA Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0892-2017Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief