FDA-Device2022-04-13Class II

BenchMark ULTRA and DISCOVERY ULTRA Instruments

Ventana Medical Systems Inc
Hazard

Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.

Sold states
US: CA, IL, PA, FL, AZ, NY, OH, CO, AR, NC, MD, AL, AK, LA, IA, MN, MA, ND, IN, TX, TN, CT, SC, GA, UT, WI, VA, SD, KY, MS, MO, DE, MI, MT, NJ, WV, OK, WA, OR, PR, DC, NM, ID, NH, NE, ME, KS, HI, NV, RI, WY, VT OUS: Australia, Austria, Belgium, Canada, China, Croatia, Denmark, Finland, France, Germany, Hong Kong, Hungary, India, Israel, Italy, Japan, Netherlands, Norway, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, United Kingdom, Algeria, Argentina, Aruba, Bosnia-Herzegovina., Brazil, Bulgaria, Chile, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Greece, Indonesia, Ireland, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Martinique, Mexico, Monaco, Morocco, Namibia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Reunion, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Vietnam, White Russia, Zimbabwe
Affected count
9214
Manufactured in
1910 E Innovation Park Dr, Oro Valley, AZ, United States
Products
BenchMark ULTRA and DISCOVERY ULTRA Instruments

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0894-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief