FDA-Device2018-03-14Class II

Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a Model 8637 SynchroMed II pump, supplied in 20 ml and 40 ml reservoir sizes, and a catheter. In the United States, the SynchroMed II Pump is specifically approved (indicated) for drugs as refer

Medtronic Neuromodulation
Hazard

Medtronic implemented a final design change intended to address the motor stall due to corrosion and is recalling devices with the previous design that are not implanted.

Sold states
US Nationwide Distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY
Affected count
27,547 US and 13,025 OUS
Manufactured in
7000 Central Ave Ne, N/A, Minneapolis, MN, United States
Products
Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a Model 8637 SynchroMed II pump, supplied in 20 ml and 40 ml reservoir sizes, and a catheter. In the United States, the SynchroMed II Pump is specifically approved (indicated) for drugs as referenced in Indications Drug Stability and Emergency Procedures Reference Manual (refer to attached labeling # MA08927A006).

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0896-2018