FDA-Device2013-03-06Class II
Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system.
Philips Medical Systems (Cleveland) Inc
Hazard
The four screws that secure the patient table top to the carrier became loose, which caused the table top to detach.
Sold states
Worldwide Distribution - USA including AL, AR, CA, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NV, NY, OH, PA, PR and VA.
Affected count
412 Units
Manufactured in
595 Miner Rd, N/A, Cleveland, OH, United States
Products
Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system.