FDA-Device2015-01-07Class II
SenSmart, Regional Oximetry EQUANOX Classic Plus Adult/Pediatric Sensor > 40 kg., Model 8003CA (REF 9824-001, 3-Wavelength), 8004CA (REF 9705-001, 4-Wavelength), with associated materials (INT-100 cable, operator's manual; and Quick Start Guide). Sensors are single-packed in a pouch, then multi-packed, 20 per shelf-box.
Nonin Medical, Inc
Hazard
Nonin Medical is conducting a recall of certain regional oximetry sensors and associated products. They have received reports of signal loss during use of our shortcabled regional oximetry sensors with diaphoretic patients. No injury or illness have been reported.
Sold states
To be provided later.
Affected count
5892 (5452 US, 440 OUS)
Manufactured in
13700 1st Ave N, N/A, Plymouth, MN, United States
Products
SenSmart, Regional Oximetry EQUANOX Classic Plus Adult/Pediatric Sensor > 40 kg., Model 8003CA (REF 9824-001, 3-Wavelength), 8004CA (REF 9705-001, 4-Wavelength), with associated materials (INT-100 cable, operator's manual; and Quick Start Guide). Sensors are single-packed in a pouch, then multi-packed, 20 per shelf-box.