FDA-Device2022-04-20Class II

Integra Universal Flexible Arm part number REF 1362275

Integra LifeSciences Corp.
Hazard

Excess force used to tighten the Universal Flexible Arm can cause the internal cable to break at the tip, resulting in an unusable condition.

Sold states
US Nationwide Distribution: AL, AZ, FL, GA, IL, MA, MD, NY, OH, OR, PA, TX
Affected count
54
Manufactured in
1100 Campus Rd, N/A, Princeton, NJ, United States
Products
Integra Universal Flexible Arm part number REF 1362275

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0906-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief