FDA-Device2020-02-12Class I

ResMed Stellar, Non-invasive/invasive ventilators, Model: 100, 150

ResMed Ltd.
Hazard

Combination of software and a component failure may cause audible alarms not to operate properly, the alarm buzzer not work, for ventilators that have a failed electronic component and, that are stored without AC power connected for more than 36 hours leading to full depletion of the battery and, that powers on automatically when connected to AC power without pressing the power switch.

Sold states
U.S.: PA, MA, VA, TN, CT, IL, GA, FL, MT
Affected count
69
Manufactured in
1 Elizabeth Macarthur Dr, N/A, Bella Vista, N/A, Australia
Products
ResMed Stellar, Non-invasive/invasive ventilators, Model: 100, 150

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0915-2020

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief