FDA-Device2013-03-13Class III
D3000 Advantage Drive System, 115v Product Usage: The Advantage Drive Systems function as a powered instrument system consisting of handpieces and accessories to perform cutting of soft tissue and bone
Linvatec Corp. dba ConMed Linvatec
Hazard
The Membrane Switch Panel (referred to as "front panel" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment: General requirements for basic safety and essential performance.
Sold states
Worldwide Distribution - USA (nationwide) including the states of: CA, FL, GA, IL, KS, MA, MD, MO, NE, NY, and TX and the countries of: Canada, Germany, Romania, Spain, Palestine, South Africa, France, Sweden and Singapore
Affected count
31 units
Manufactured in
11311 Concept Blvd, N/A, Largo, FL, United States
Products
D3000 Advantage Drive System, 115v Product Usage: The Advantage Drive Systems function as a powered instrument system consisting of handpieces and accessories to perform cutting of soft tissue and bone