FDA-Device2021-02-03Class II
Philips Respironics V60 Ventilator Part Number DU1053617 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. Philips Respironics V60 Ventilator Part Number U1053617 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. Note: U1053617 added 1/29/2021 after noticing
Respironics California, LLC
Hazard
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments.
Sold states
Worldwide distribution - US Nationwide including Guam and Puerto Rico; Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia.
Affected count
481 systems (DU1053617)(Amended 07/29/2021 - 44,666 U.S. systems for faceplate replacement), and 292 systems (U1053617) (Amended 07/29/2021 - 7,770)
Manufactured in
2271 Cosmos Ct, Carlsbad, CA, United States
Products
Philips Respironics V60 Ventilator Part Number DU1053617 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. Philips Respironics V60 Ventilator Part Number U1053617 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. Note: U1053617 added 1/29/2021 after noticing it had been inadvertently left off
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0916-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief